Relative detection efficiency of back- and front-illuminated charge-coupled device cameras for x-rays between 1 keV and 18 keV
-
Szlachetko, Jakub
Department of Physics, University of Fribourg, Switzerland
-
Dousse, Jean-Claude
Department of Physics, University of Fribourg, Switzerland
-
Hoszowska, Joanna
Department of Physics, University of Fribourg, Switzerland
-
Berset, Michel
Department of Physics, University of Fribourg, Switzerland
-
Cao, Wei
Department of Physics, University of Fribourg, Switzerland
-
Szlachetko, Monika
Department of Physics, University of Fribourg, Switzerland
-
Kavčič, Matjaz
J. Stefan Institute, Ljubljana, Slovenia
Show more…
Published in:
- Review of Scientific Instruments. - 2007, vol. 78, no. 9, p. 093102
English
High-resolution x-ray measurements were performed with a von Hamos-type bent crystal spectrometer using for the detection of the diffracted photons either a back- illuminated charge-coupled device (CCD) camera or a front-illuminated one. For each CCD the main x-ray emission lines (e.g., Kα, Kβ, Lα, and Lβ) of a variety of elements were measured in order to probe the performances of the two detectors between 1 and 18 keV. From the observed x-ray lines the linearity of the energy response, the noise level, the energy resolution, and the quantum efficiency ratio of the two CCDs were determined.
-
Faculty
- Faculté des sciences et de médecine
-
Department
- Département de Physique
-
Language
-
-
Classification
-
Physics
-
License
-
License undefined
-
Identifiers
-
-
Persistent URL
-
https://folia.unifr.ch/unifr/documents/300628
Statistics
Document views: 49