Journal article
Determination of 137Cs half-life with an ionization chamber.
-
Juget F
Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, CH-1007 Lausanne, Switzerland. Electronic address: Frederic.juget@chuv.ch.
-
Nedjadi Y
Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, CH-1007 Lausanne, Switzerland.
-
Buchillier T
Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, CH-1007 Lausanne, Switzerland.
-
Bochud F
Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, CH-1007 Lausanne, Switzerland.
-
Bailat C
Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, CH-1007 Lausanne, Switzerland.
Show more…
Published in:
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. - 2016
English
The half-life of 137Cs was measured with an ionization chamber by following the decay of 5 sources over a 30 years period between 1983 and 2013. The ratio between the ionization chamber current for the cesium sources and 226Ra source was used for the half-life calculation. The value found for the 137Cs half-life is 10,955.2±10.7 days, where the uncertainty evaluation combines type A and B for one standard deviation.
-
Language
-
-
Open access status
-
closed
-
Identifiers
-
-
Persistent URL
-
https://folia.unifr.ch/global/documents/281365
Statistics
Document views: 23