Influence of flexible fixation for open book injury after pelvic trauma--a biomechanical study.
Journal article

Influence of flexible fixation for open book injury after pelvic trauma--a biomechanical study.

  • Stuby FM BG Trauma Center, Eberhard Karls University, Schnarrenbergstrasse 95, D-72076 Tuebingen, Germany. Electronic address: fabian.stuby@uni-tuebingen.de.
  • Doebele S BG Trauma Center, Eberhard Karls University, Schnarrenbergstrasse 95, D-72076 Tuebingen, Germany. Electronic address: sdoebele@bgu-tuebingen.de.
  • Agarwal Y AO Research Institute Davos, Clavadelerstrasse 8, CH-7270 Davos, Switzerland. Electronic address: yash.agarwal@aofoundation.org.
  • Windolf M AO Research Institute Davos, Clavadelerstrasse 8, CH-7270 Davos, Switzerland. Electronic address: markus.windolf@aofoundation.org.
  • Gueorguiev B AO Research Institute Davos, Clavadelerstrasse 8, CH-7270 Davos, Switzerland. Electronic address: boyko.gueorguiev@aofoundation.org.
  • Ochs BG BG Trauma Center, Eberhard Karls University, Schnarrenbergstrasse 95, D-72076 Tuebingen, Germany. Electronic address: gochs@bgu-tuebingen.de.
Show more…
  • 2014-05-24
Published in:
  • Clinical biomechanics (Bristol, Avon). - 2014
English BACKGROUND
Implant loosening is frequently detected after fixation of open book injuries. Though many authors do not see this as a complication, it is often the reason for hardware removal or reinstrumentation in the case of remaining instability. We hypothesized that the flexibility of the implant has an influence on loosening and thus on failure of the construct.


METHODS
We used 6 fresh-frozen pelvic specimens and tested them with our recently introduced test setup for two-leg alternate loading. We subjected them to a non-destructive quasi-static test in the intact condition followed by a non-destructive cyclic test under axial sinusoidal loading with progressive amplitude. Afterwards we simulated an open book injury and performed fixation with three different configurations of a modular fixation system (1-, 2- or 4-rod configuration) in randomized order. Subsequently, the specimens were subjected to 3 cyclic tests with the same loading protocol as previously defined. Finally, each construct was cyclically tested to failure keeping the final rod configuration.


FINDINGS
We detected significantly greater mobility after 1-rod-fixation and no significant differences after 2-rod or 4-rod-fixation compared to the intact symphysis condition. In the destructive test series the 4-rod-fixation failed first followed by the 1-rod-fixation. The 2-rod-fixation sustained almost 3 times as many load cycles prior to failure as the 4-rod-fixation, whereas the 1-rod-fixation sustained twice as many cycles as the 4-rod-fixation.


INTERPRETATION
In conclusion, flexible fixation of the ruptured pubic symphysis in human specimens shows superior behavior with respect to load bearing capacity and ability to withstand cyclic loading compared to stiff constructs.