  <dc:creator>Zeeshan, Faisal</dc:creator>
  <dc:creator>Hoszowska, Joanna</dc:creator>
  <dc:creator>Loperetti-Tornay, Lucie</dc:creator>
  <dc:creator>Dousse, Jean-Claude</dc:creator>
  <dc:date>2019</dc:date>
  <dc:description xmlns:ns0="xml" ns0:lang="en">We report on a laboratory-based facility for in-house x-ray absorption fine structure  (XAFS) measurements. The device consists of a conventional x-ray source for the  production of the incident polychromatic radiation and a von Hamos bent crystal  spectrometer for the analysis of the incoming and transmitted radiation. The reliability  of the laboratory-based setup was evaluated by comparing the Cu K-edge and Ta L3- edge XAFS spectra obtained in-house with the corresponding spectra measured at a  synchrotron radiation facility. To check the accuracy of the device, the K- and L-edge  energies and the attenuation coefficients below and above the edges of several 3d,  4d, and 5d elements were determined and compared with the existing experimental  and theoretical data. The dependence of the XAFS spectrum shape on the oxidation  state of the sample was also probed by measuring inhouse the absorption spectra of  metallic Fe and two Fe oxides (Fe2O3 and Fe3O4).</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:identifier>https://folia.unifr.ch/unifr/documents/308125</dc:identifier>
  <dc:identifier>https://folia.unifr.ch/documents/308125/files/dou_isl.pdf</dc:identifier>
  <dc:language>eng</dc:language>
  <dc:relation>info:eu-repo/semantics/altIdentifier/doi/10.1063/1.5094873</dc:relation>
  <dc:rights>info:eu-repo/semantics/openAccess</dc:rights>
  <dc:rights>License undefined</dc:rights>
  <dc:source>Review of Scientific Instruments. - 2019, vol. 90, no. 7, p. 073105</dc:source>
  <dc:subject>info:eu-repo/classification/udc/53</dc:subject>
  <dc:title xmlns:ns1="xml" ns1:lang="en">In-house setup for laboratory-based x-ray absorption fine structure spectroscopy measurements</dc:title>
  <dc:type>http://purl.org/coar/resource_type/c_6501</dc:type>
</oai_dc:dc>